Dental Vacuum System
Part Numbers: VS2O, VS40, VS50, VS50H, VS80 and VS80H


OPERATING INFORMATION
AT THE START OF THE DAY:
Always TURN ON THE WATER before TURNING ON THE POWER .
The VacStar may be turned on/off from a single, convenient location within the dental suite using a Remote Control Panel (See Optional Accessories) .
The vacuum level is factory preset at 10 In Hg (inches of mercury) . This is the reading on the gauge when all HVE’s (High Volume Evacuator) and SE’s (Saliva Ejector) are CLOSED . Should this setting be too high for your needs, contact your dealer to readjust the setting .
It is recommended that the system run continuously during the day . However, the VacStar can be turned off if suction is not required for a period of 15 minutes or longer .
If one pump is being operated at a time, it is important to alternate pumps on an every other day schedule so that the pumps are used evenly .
AT THE END OF THE DAY: Always TURN THE POWER OFF, then TURN THE WATER OFF